Aims
Sacral neuromodulation (SNM) is a guideline-recommended treatment with proven therapeutic benefit for urinary urgency incontinence (UUI) patients. The Axonics® System is the first Food and Drug Administration-approved rechargeable SNM system and is designed to deliver therapy for a minimum of 15 years. The ARTISAN-SNM study was designed to evaluate UUI participants treated with the Axonics System. Two-year follow-up results are presented.Methods
One hundred and twenty-nine UUI participants underwent implantation with the Axonics System. Therapeutic response rate, participant quality of life (QoL), and satisfaction were determined using 3-day voiding diaries, ICIQ-OABqol, and satisfaction questionnaires. Participants were considered responders if they had a 50% or greater reduction in UUI episodes post-treatment. As-treated and Completers analyses are presented.Results
At 2 years, 93% of the participants (n = 121 Completers at 2 years) were therapy responders, of which 82% achieved ≥ 75% reduction in UUI episodes and 37% were dry (100% reduction). Daily UUI episodes reduced from 5.6 ± 0.3 at baseline to 1.0 ± 0.2 at 2 years. Statistically significant improvements in ICIQ-OABqol were reported. All participants were able to recharge their device and 94% of participants reported that the recharging frequency and duration were acceptable. Participant demographics nor condition severity were correlated with clinical outcomes or recharging experience. No unanticipated or serious device-related adverse events occurred.Conclusions
At 2 years, participants treated with the Axonics System demonstrated sustained safety and efficacy, high levels of satisfaction with therapy and recharging. Participant-related factors were not associated with efficacy or recharging outcomes, indicating the reported results are applicable to a diverse population.